RIC at Ljubljana, Slovenia (Region VI)
|
(Information on this webpage
is based on the latest report from the RIC.
Please refer to the RIC's website for the latest information.) |
|
General Information |
Address:
Slovenian Environment Agency,
Vojkova 1b, 1000 Ljubljana, Slovenia |
Website:
http://www.arso.gov.si/en/about%20the%20agency/laboratories/ |
Contact person: Drago GROSELJ |
Email:
Drago.Groselj@gov.si |
Tel: +386 1 478 4100 |
Fax: +386 1 478 4052 |
Calibration capabilities: Temperature, Relative Humidity, Atmospheric Pressure, Air Quality |
|
Specific Information on instrument calibration capabilities |
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Platinum resistance thermometers |
0.01 °C |
Triple point of water cell, Isotech, 811L-32 |
0.7 mK |
7 June 2015 |
LMK (Slovenia) |
Platinum resistance thermometers |
29.7646 °C |
Gallium melting point cell, Isotech, ITL-M-17401 |
0.9 mK |
7 June 2015 |
LMK (Slovenia) |
Resistance thermometers |
-50 to 50 °C |
Standard platinum resistance thermometer (Isotech SPRT 419, Fluke 7381, DC bridge MI6010) |
12 mK |
6 June 2018 |
LMK (Slovenia) |
Liquid-in-glass themrometers |
-20 to 50 °C |
Platinum resistance thermometer (Fluke 5626, Fluke 7381, Fluke 1595A) |
25 mK |
20 August 2018 |
Slovenian Environment Agency |
Self-indicated thermometers |
-50 to 50 °C |
Platinum resistance thermometer (Fluke 5626, Fluke 7381, Fluke 1595A)
|
25 mK |
20 August 2018 |
Slovenian Environment Agency |
|
Status of accreditation (date of the latest accreditation): 11 June 2018 |
Link to the accreditation certificate |
Accreditation body: Slovenian Accreditation (SA) |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Hygrometer |
10 to 95 %RH |
Dew-point hygrometer
(MBW 373LHX,Thunder Scientific 2500ST) |
1.9 %RH (Ta=-10 °C to 0 °C)
1.6 %RH (Ta=0 °C to 10 °C)
1.0 %RH (Ta=10 °C to 50 °C)
|
14 November 2018 |
LMK (Slovenia) |
Hygrometer |
10 to 95 %RH |
Dew-point meter
(MBW 473, Vötsch VC3 7100) |
4.2 %RH (Ta=-20 °C to 0 °C)
3.2 %RH (Ta=0 °C to 40 °C) |
7 September 2018 |
LMK (Slovenia) |
|
Status of accreditation (date of the latest accreditation): 11 June 2018 |
Link to the accreditation certificate |
Accreditation body:
Slovenian Accreditation (SA)
|
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Digital barometer |
200 to 1800 hPa |
Digital piston gauge
Ruska 2465A-754
(Generator DH type PPC1) |
3.1 Pa + 1.9*10-5 P |
26 May 2017 |
CMI (Czech Republic) |
Barometer without tube connection |
800 to 1050 hPa |
Fluke RPM4 (barometric chamber Kambic VTK-02) |
65 Pa |
22 May 2018 |
Slovenian Environment Agency |
|
Status of accreditation (date of the latest accreditation): 11 June 2018 |
Link to the accreditation certificate |
Accreditation body: Slovenian Accreditation (SA) |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Cylinders or generators CO |
300 to 15000 ppbv |
Reference analyser, CRMs |
240 ppbv + 0.028 • c(CO) |
15 November 2018 |
CHMI (Czech Republic) |
Cylinders or generators SO2 |
3 to 500 ppbv |
Reference analyser, CRMs |
2.3 ppbv + 0.055 • c(SO2) |
15 November 2018 |
CHMI (Czech Republic) |
Cylinders or generators NOx |
2 to 500 ppbv |
Reference analyser, CRMs |
1.9 ppbv + 0.063 • c(NO) |
15 November 2018 |
CHMI (Czech Republic) |
Cylinders or generators NO2 |
4 to 500 ppbv |
Reference analyser, CRMs |
3.7 ppbv + 0.062 • c(NO2) |
15 November 2018 |
CHMI (Czech Republic) |
Generators O3 |
6 to 500 ppbv |
Generator/analyser O3 |
5.0 ppbv + 0.034 • c(O3) |
15 November 2018 |
CHMI (Czech Republic) |
Analyser CO |
0 to 13700 ppbv |
CRMs, zero air generator |
170 ppbv + 0.02 • c(CO) |
15 November 2018 |
CHMI (Czech Republic) |
Analyser SO2 |
0 to 380 ppbv |
CRMs, zero air generator |
1.8 ppbv + 0.05 • c(SO2) |
15 November 2018 |
CHMI (Czech Republic) |
Analyser NO (NOx) |
0 to 427 ppbv |
CRMs, zero air generator |
1.5 ppbv + 0.047 • c(NO) |
15 November 2018 |
CHMI (Czech Republic) |
Analyser O3 |
0 to 500 ppbv |
CRMs, zero air generator |
4.9 ppbv + 0.03 • c(O3) |
15 November 2018 |
CHMI (Czech Republic) |
|
Status of accreditation (date of the latest accreditation): 11 June 2018 |
Link to the accreditation certificate |
Accreditation body: Slovenian Accreditation (SA) |
|