WWW > IMOP > Regional Instrument Centresl
 

RIC at Ljubljana, Slovenia (Region VI)

(Information on this webpage is based on the latest report from the RIC.
Please refer to the RIC's website for the latest information.)
 
General Information
Address: Slovenian Environment Agency, Vojkova 1b, 1000 Ljubljana, Slovenia
Website: http://www.arso.gov.si/en/about%20the%20agency/laboratories/
Contact person: Drago GROSELJ
Email: Drago.Groselj@gov.si
Tel: +386 1 478 4100
Fax: +386 1 478 4052
Calibration capabilities: Temperature, Relative Humidity, Atmospheric Pressure, Air Quality
 
Specific Information on instrument calibration capabilities
Temperature
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Platinum resistance thermometers

0.01 °C

Triple point of water cell, Isotech, 811L-32

0.7 mK

7 June 2015

LMK (Slovenia)

Platinum resistance thermometers

29.7646 °C

Gallium melting point cell, Isotech, ITL-M-17401

0.9 mK

7 June 2015

LMK (Slovenia)

Resistance thermometers

-50 to 50 °C

Standard platinum resistance thermometer (Isotech SPRT 419, Fluke 7381, DC bridge MI6010)

12 mK

6 June 2018

LMK (Slovenia)

Liquid-in-glass themrometers

-20 to 50 °C

Platinum resistance thermometer  (Fluke 5626, Fluke 7381, Fluke 1595A)

25 mK

20 August 2018

Slovenian Environment Agency

Self-indicated thermometers

-50 to 50 °C

Platinum resistance thermometer (Fluke 5626, Fluke 7381, Fluke 1595A)

25 mK

20 August 2018

Slovenian Environment Agency

Status of accreditation (date of the latest accreditation): 11 June 2018
Link to the accreditation certificate
Accreditation body:  Slovenian Accreditation (SA)
 
Relative Humidity
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Hygrometer

10 to 95 %RH

Dew-point hygrometer

(MBW 373LHX,Thunder Scientific 2500ST)

1.9 %RH (Ta=-10 °C to 0 °C)

1.6 %RH (Ta=0 °C to 10 °C)

1.0 %RH (Ta=10 °C to 50 °C)

14 November 2018

LMK (Slovenia)

Hygrometer

10 to 95 %RH

Dew-point meter

(MBW 473, Vötsch VC3 7100)

4.2 %RH (Ta=-20 °C to 0 °C)

3.2 %RH (Ta=0 °C to 40 °C)

7 September 2018

LMK (Slovenia)

Status of accreditation (date of the latest accreditation): 11 June 2018
Link to the accreditation certificate
Accreditation body: Slovenian Accreditation (SA)
 
Atmospheric Pressure
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Digital barometer

200 to 1800 hPa

   Digital piston gauge

Ruska 2465A-754

(Generator DH type PPC1)

3.1 Pa + 1.9*10-5 P

26 May 2017

CMI (Czech Republic)

Barometer without tube connection

800 to 1050 hPa

Fluke RPM4 (barometric chamber Kambic VTK-02)

65 Pa

22 May 2018

Slovenian Environment Agency

Status of accreditation (date of the latest accreditation): 11 June 2018
Link to the accreditation certificate
Accreditation body: Slovenian Accreditation (SA)
 
Air Quality
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Cylinders or generators CO

300 to 15000 ppbv

Reference analyser, CRMs

240 ppbv + 0.028 • c(CO)

15 November 2018

CHMI (Czech Republic)

Cylinders or generators SO2

3 to 500 ppbv

Reference analyser, CRMs

2.3 ppbv + 0.055 • c(SO2)

15 November 2018

CHMI (Czech Republic)

Cylinders or generators NOx

2 to 500 ppbv

Reference analyser, CRMs

1.9 ppbv + 0.063 • c(NO)

15 November 2018

CHMI (Czech Republic)

Cylinders or generators NO2

4 to 500 ppbv

Reference analyser, CRMs

3.7 ppbv + 0.062 • c(NO2)

15 November 2018

CHMI (Czech Republic)

Generators O3

6 to 500 ppbv

Generator/analyser O3

5.0 ppbv + 0.034 • c(O3)

15 November 2018

CHMI (Czech Republic)

Analyser CO

0 to 13700 ppbv

CRMs, zero air generator

170 ppbv + 0.02 • c(CO)

15 November 2018

CHMI (Czech Republic)

Analyser SO2

0 to 380 ppbv

CRMs, zero air generator

1.8 ppbv + 0.05 • c(SO2)

15 November 2018

CHMI (Czech Republic)

Analyser NO (NOx)

0 to 427 ppbv

CRMs, zero air generator

1.5 ppbv + 0.047 • c(NO)

15 November 2018

CHMI (Czech Republic)

Analyser O3

0 to 500 ppbv

CRMs, zero air generator

4.9 ppbv + 0.03  • c(O3)

15 November 2018

CHMI (Czech Republic)

Status of accreditation (date of the latest accreditation): 11 June 2018
Link to the accreditation certificate
Accreditation body: Slovenian Accreditation (SA)
 

* A CMC (calibration and measurement capability) is the smallest uncertainty (k=2) of measurement that can be expected to be achieved by the RIC during a calibration under normal conditions. This CMC is evaluated by the RIC itself and described in the scope of accreditation of the RIC, if available.

 
Further Information
 
 
 



top

Top






©2017 World Meteorological Organization
, 7bis, avenue de la Paix, Case postale No. 2300, CH-1211 Geneva 2, Switzerland 
Tel.: +41(0)22 730 81 11 - Fax: +41(0)22 730 81 81
Disclaimer | UN System | Accessibility

 

  Extranet Homepage Website Public Comms