Calibration capabilities: Temperature, Relative Humidity, Atmospheric Pressure, Wind, Precipitation, Air Quality |
|
Specific Information on instrument calibration capabilities |
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
In-Glass thermometers |
-30 to 40 °C |
In-glass thermometer |
0.2 °C |
2019 |
SMU |
Platinum resistance thermometers |
-30 to 40 °C |
SPRT + resistance bridge |
0.05 °C |
2018 |
SMU |
Electronic thermometers |
-30 to 40 °C |
SPRT + resistance bridge |
0.05 °C |
2018 |
SHMU |
Electronic thermometers |
-30 to 40 °C |
PRT + data collector |
0.3 °C |
2019 |
SHMU |
Bimetal thermometers |
-30 to 40 °C |
PRT + data collector |
1 °C |
2019 |
SHMU |
|
Status of accreditation (date of the latest accreditation): 17 September 2018 |
Link to the Certificate of Accreditation |
Accreditation body: Slovak National Accreditation Service (SNAS) |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Electronic RH sensors |
10 to 97 %RH |
Dew-point standard |
2.5 %RH |
2018 |
CMI |
Electronic RH sensors |
20 to 95 %RH |
Capacitance RH sensor in climatic chamber |
3 %RH |
2018 |
SHMU |
Mechanical hygrometers |
20 to 95 %RH |
Capacitance RH sensor in climatic chamber |
3 %RH |
2018 |
SHMU |
|
Status of accreditation (date of the latest accreditation): 17 September 2018 |
Link to the Certificate of Accreditation |
Accreditation body:
Slovak National Accreditation Service (SNAS) |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Absolute pressure |
700 to 1050 hPa |
Calibrator DHI PPC2+
(laboratory calibration) |
0.1 hPa |
2019 |
SMU |
Absolute pressure |
700 to 1050 hPa |
Barometer Paroscientific model 760
(field calibration) |
0.15 hPa |
2019 |
SMU |
|
Status of accreditation (date of the latest accreditation): 17 September 2018 |
Link to the Certificate of Accreditation |
Accreditation body: Slovak National Accreditation Service (SNAS) |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Anemometer |
0.3 to 40 m/s |
Thermal and rotating anemometer,
Pitot tube |
(0.13*v+0.1) m/s |
2018 |
CMI |
|
Status of accreditation (date of latest accreditation): 17 September 2018 |
Link to the Certificate of Accreditation |
Accreditation body: Slovak National Accreditation Service (SNAS) |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Tipping bucket |
10 to 25 mm |
Volumetric flask |
2 % |
2005 |
SMU |
Weight rain gauge |
0.1 to 260 mm |
Set of weights |
0.02 mm |
2017 |
SLM |
|
Status of accreditation (date of latest accreditation): 17 September 2018 |
Link to the Certificate of Accreditation |
Accreditation body: Slovak National Accreditation Service (SNAS) |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
NOx analyzers |
0.5 to 500 nmol/mol |
Standard analyzer |
(0.04 hm + 2) nmol/mol |
2017 |
CHMI |
SO2 analyzers |
0.1 to 500 nmol/mol |
Standard analyzer |
(0.05 hm + 2) nmol/mol |
2017 |
CHMI |
O3 analyzers |
0.5 to 500 nmol/mol |
Standard analyzer |
(0.03 hm + 1.5) nmol/mol |
2017 |
CHMI |
CO analyzers |
0.01 to 20 μmol/mol |
Standard analyzer |
(0.02 hm + 0.4) μmol/mol |
2017 |
CHMI |
|
Status of accreditation (date of latest accreditation): 17 September 2018 |
Link to the Certificate of Accreditation |
Accreditation body: Slovak National Accreditation Service (SNAS) |
|