RIC at Casablanca, Morocco (Region I) |
(Information on this webpage is based on the latest report from the RIC.
Please refer to the RIC's website for the latest information.) |
|
General Information |
Address: Direction de la Météorologie Nationale,
Avenue Capitane Tayeb Naciri, administrative complex Hay Hassani, 20220 Casablanca, Morocco |
Website: http://cri.marocmeteo.ma/ |
Contact person: Mounir AZIZ |
Email: azizmounirgmail.com |
Tel: +212 661 47 23 98 |
Fax:
+212 522 90 85 93 |
Calibration capabilities: Temperature, Relative Humidity, Atmospheric Pressure, Precipitation, Solar Radiation |
|
Specific Information on instrument calibration capabilities |
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Resistance thermometer PT100 |
-20 to 60 °C | SPRT 100 (ISOTECH) |
0.065 °C |
15 February 2017 |
Laboratory AIRMETROLOGY Cofrac accreditation n°02-60-18 |
Triple point of water cell |
Gallium fixed point cell |
Liquid in Glass thermometer |
MicroK precision thermometry bridge |
|
Status of accreditation (date of the latest accreditation): |
Link to the accreditation certificate |
Accreditation body: |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Capacitive hygrometer |
|
Dew Point Mirror type MBW473 |
1.6 %RH |
14 February 2017 |
Laboratory AIRMETROLOGY Cofrac accreditation n°02-60-19 |
Display hygrometer |
Hygrogen generator |
Hygrograph |
Climatic chamber |
|
Status of accreditation (date of the latest accreditation): |
Link to the accreditation certificate |
Accreditation body: |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Electric/Digital barometers |
500 to 1100 hPa (absolute pressure) |
Barometer CPC6000 IS50; Nitrogen Generator CPC600
|
0.015 hPa + 3.3*10-5 P |
15 September 2018 |
MENSOR ZL accreditation N:2066.01 |
Pressure gauge
|
1 to 25 bars (gauge pressure)
|
Manometer CPC6000 IS-50; Nitrogen Generator CPC 6000
|
14.1 + 3.9*10-5 P [Pa]
|
|
Status of accreditation (date of the latest accreditation): |
Link to the accreditation certificate |
Accreditation body: |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Tipping bucket rain gauge (0.1; 0.2; 0.5 mm resolution) |
0 to 10 mm/minute |
Pluvio Precipitation Weighing Gauge
|
0.3 % |
21 May 2018 |
National Metrology Lab |
|
|
Status of accreditation (date of the latest accreditation): |
Link to the accreditation certificate |
Accreditation body: |
|
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Pyranometer |
up to 1400 W/m² | Kipp & Zonnen Calibration bench (with projector) |
1 % |
August 2018 |
Kipp&Zonen |
Pyrheliometer |
|
Pyrheliometer |
|
Status of accreditation (date of the latest accreditation): |
Link to the accreditation certificate |
Accreditation body: |
|
|