RIC at Tsukuba, Japan (Region II) |
(Information on this webpage
is based on the latest report from the RIC.
Please refer to the RIC's website for the latest information.) |
| |
General Information |
| Address: 1-2 Nagamine Tsukuba Ibaraki 305-0052, Japan |
| Website: http://www.jma.go.jp/jma/jma-eng/jma-center/ric/RIC_HP.html |
| Contact person: Kouichi
NAKASHIMA |
Email:
kouichi.nakashima met.kishou.go.jp; ric-tsukuba met.kishou.go.jp |
| Tel: +81 298 51 4123 |
| Fax:+81 298 51 1670 |
| Calibration capabilities: Temperature, Relative Humidity, Atmospheric Pressure, Wind |
| |
Specific Information on instrument calibration capabilities |
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Contact Type Thermometer |
|
Platinum resistance thermometer NSR-160 (Netsushin) |
-40 to < 0 °C: 45 mK,
at 0 °C: 13 mK,
0 < to 50 °C: 36 mK
|
27 August 2018 |
Tanaka Kikinzoku Kogyo K.K. Isehara Works Thermometer Calibration Laboratory |
|
| Status of accreditation (date of the latest accreditation): 29 August 2017 |
| Link to the accreditation certificate |
| Accreditation body: National Institute of Technology and Evaluation (Japan) |
| |
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Dew-point hygrometer |
Dew point:
-5 to 0 °C |
Chilled-mirror dew-point hygrometer (display)
DewStar S-1M-0 (sensor)
DewStar S-2S-0K (Shinyei technology, Japan)
|
Dew point 0.12 °C
|
14 February 2018 |
National Metrology Institute of Japan |
|
Dew point 0.09 °C |
Electronic hygrometer |
20 to < 30 %RH
(for temp: 20 to 26 °C)
|
0.8 %RH |
30 to 95 %RH
(for temp: 20 to 26 °C) |
1.7 %RH |
20 to < 30 %RH
(for temp: 20 - 26 °C)(1)
|
0.6 %RH |
30 to 95 %RH
(for temp: 20 - 26 °C)(1) |
1.4 %RH |
|
| (1) - Calibration which regards a dew-point hygrometer as the hygrometer of a relative humidity indication |
| Status of accreditation (date of the latest accreditation): 29 August 2017 |
| Link to the accreditation certificate |
| Accreditation body: National Institute of Technology and Evaluation (Japan) |
| |
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Digital pressure gauge
|
|
Pressure balance AV-02 (Futaba sokki)
|
The larger of:
0.0085 % or 7.5 Pa
|
19 September 2018 |
National Metrology Institute of Japan |
|
| Status of accreditation (date of the latest accreditation): 29 August 2017 |
| Link to the accreditation certificate |
| Accreditation body: National Institute of Technology and Evaluation (Japan) |
| |
|
Instrument Undergoing Calibration |
Calibration Range |
Reference standard, Equipment |
Calibration and Measurement Capability (CMC)* |
Traceability of Reference equipment |
Last standard calibration date |
Calibration body |
Anemometer |
0.5 to 90 m/s |
From 0 to 20 m/s:
Ultrasonic anemometer
DA-700 (Sonic, Japan)
|
N/A |
1
December 2017 |
National Metrology Institute of Japan |
From 20 to 90 m/s:
Pitot tube JB1512154 (Tsukuba Rikaseiki, Japan)
Differential pressure gauge MT210 (2 sets)
(Yokogawa, Japan) |
5 March 2018 |
|
| Status of accreditation: |
| Link to the accreditation certificate |
| Accreditation body: |