General Information
Address: Meteorological Observation Centre of China Meteorological Administration, No. 46, Zhongguancun Nandajie, 100081 Beijing, China
Website:
Contact person: Xuejing NAN
Email: 576809156@qq.com
Tel: 8610-68400493
Fax: 8610-68409767
Calibration capabilities: Temperature, Relative Humidity, Atmospheric Pressure, Wind, Precipitation, Solar Radiation
 
Specific Information on instrument calibration capabilities
Temperature
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Standard Mercury-in-Glass Thermometers;

Pt Resistance Thermometers

-60 to 80 °C

First-class Standard Pt Resistance Thermometer

50 mK

29 March 2018

National Institute of Metrology, China

Status of accreditation (date of the latest accreditation): 27 June 2018
Link to the accreditation certificate: Registration No. L11135
Accreditation body: China National Accreditation Service for Conformity Assessment (CNAS)
 
Relative Humidity
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Standard Ventilation Psychrometers;

Hygrometers

5 to 98 %RH

Precision Dew-point Hygrometer

0.4 to 1.5 %RH

12 April 2018

National Institute of Metrology, China

Status of accreditation (date of the latest accreditation): 27 June 2018
Link to the accreditation certificate: Registration No. L11135
Accreditation body: China National Accreditation Service for Conformity Assessment (CNAS)
 
Atmospheric Pressure
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Digital barometers

90 to 3500 hPa

   Gas Piston Pressure Gauge

1 to 4 Pa

(100 to 1200 hPa)

26 May 2017

National Institute of Metrology, China

Status of accreditation (date of the latest accreditation): 27 June 2018
Link to the accreditation certificate: Registration No. L11135
Accreditation body: China National Accreditation Service for Conformity Assessment (CNAS)
 
Wind
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body
Portable 3-cup anemometers

0.2 to 70 m/s

Pilot tube

0.07 m/s

(2 to 30 m/s)

7 April 2014

National Institute of Metrology, China

Wind direction and wind speed sensor for automatic weather station

0.08 m/s

(2 to 40 m/s)

0.08 to 0.18 m/s

(40 to 60 m/s)

Magnetoelectricity wind speed sensor for wind farm

0.07 m/s

(3 to 30 m/s)

Hot bulb anemometer

0.07 to 0.25 m/s

(0.5 to 2 m/s)

0.07 m/s

(2 to 30 m/s)

Status of accreditation (date of the latest accreditation): 27 June 2018
Link to the accreditation certificate: Registration No. L11135
Accreditation body: China National Accreditation Service for Conformity Assessment (CNAS)
 
Precipitation
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Raingauges, Measuring cylinders

0 to 942.48 mL

Standard Capacity Measurement (glass)

MPE: ± (0.031 to 0.314) mL

October 2016

National Institute of Metrology, China

Status of accreditation (date of the latest accreditation):
Link to the accreditation certificate
Accreditation body:
 
Solar Radiation
Instrument Undergoing Calibration
Calibration Range
Reference standard, Equipment
Calibration and Measurement Capability (CMC)*
Traceability of Reference equipment
Last standard calibration date
Calibration body

Pyrheliometer

5 to 15 μV/(W/m²)

Cavity Radiometer

0.6 %

October 2015

PMOD-WRC Davos, Switzerland

Pyranometer

5 to 25 μV/(W/m²)

Secondary Standard Pyranometer

1.6 %

November 2018

National Center for Meteorological Metrology,China

Status of accreditation (date of the latest accreditation): 27 June 2018
Link to the accreditation certificate: Registration No. L11135
Accreditation body: China National Accreditation Service for Conformity Assessment (CNAS)
 

* A CMC (calibration and measurement capability) is the smallest uncertainty (k=2) of measurement that can be expected to be achieved by the RIC during a calibration under normal conditions. This CMC is evaluated by the RIC itself and described in the scope of accreditation of the RIC, if available.

 